We use cookies to enhance your browsing experience, serve personalized ads or content, and analyze our traffic. By clicking "Accept All", you consent to our use of cookies. Read more about privacy policy.

7 - Spectrophotometer

Sold

7 - Spectrophotometer

Location:

Description

Technical data: Photometer: Measuring principle: Absorption single beam photometer with reference beam and several fixed wavelenghts. Light source: Xenon flash light. Monochromator: Holographic concave grating. Beam receiver: Silicon photo diodes. Wavelengths: 230, 260, 280, 340, 405, 490, 550, 595, 650 nm. Wavelength selection: Dependent on method, controlled by program. Spectral bandwidth 5 nm at 230 to 340 nm, 7 nm at 405 to 650 nm. Photometric measuring range Quartz glass cuvette: • A = 0 to 3, except: • A = 0 to 2 at 340 nm, • Only for dye methods: A = 0 to 2 at 550/650 nm, UVette (Eppendorf): • A = 0 to 2.5 at 230 nm, • A = 0 to 2.6 at 260 nm, • A = 0 to 2.8 at 280 nm, • Other values see quartz glass cuvette, Reading accuracy ΔA = 0.001, Random photometric error ≤ 0.002 at A = 0, ≤ 0.005 at A = 1. Systematic photometric error ± 1% at A = 1. Stray light component < 0.05 %. Power supply: Power supply 100 to 240 V ± 10 % / 50 to 60 Hz ± 5 %. Overvoltage category IEC 61010-1 category II. Degree of contamination: IEC 61010-1 category II. Power consumption approx. 20 W Weight / dimensions: Weight 3 kg. Dimensions Width: 200 mm, Depth: 320 mm, Height: 100 mm, Space required Width: 400 mm (or 650 mm including printer), Depth: 500 mm. Interfaces: Interface for printer and PC serial RS-232

Specifications

ManufacturerEppendorf
ModelBioPhotometer plus 6132
ConditionUsed
Serial Number6132 01465
Online Auctions2023-06-30 16:00:00